Technical Specifications OMNI Express OMNI K OMNI OMNI II OMNI III
Display 7” color TFT 10.5” touch screen 10.1” touch screen 12” touch screen 15” touch screen
Trace 2 or 3 waveforms 8 waveforms
Indicator Alarm Indicator
Power Indicator
QRS beep and alarm sound
Trend Time From 30 minutes to 72 hours 1 – 72 hour
ECG
Input 5-lead or 3 lead ECG cable and standard AAMI line for connection 5-lead ECG cable and standard AAMI line for connection
Lead Choice I, II, III, aVR, aVF, aVL, V I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
Gain Choice X0.5, x1.0, x2.0 X0.5, x1, x2, x4
Frequency Characteristic 0.67 ~ 40 HZ (+3dB attenuation) 0.05 ~ 35 HZ (+3dB)
ECG Waveforms 7 channels
Sweep Speed 12.5, 25, 50 mm/s 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
HR Display Range 30 ~ 300bpm
Accuracy ±1bpm or ±1%, whichever is greater
Alarm Limit Range Setting Upper limit 80~400bpm
Lower limit 20~150bpm
Upper limit 100~200bpm
Lower limit 30~100bpm
RESP
Measure Method RA-LL impedance
Range 0 ~ 120 rpm
Accuracy ±3 rpm
Alarm Limit Setting Upper limit 6~120 rpm
Lower limit 3~120 rpm
Sweep Speed 12.5 and 25mm/s 12.5, 25, 50 and 100 mm/sec
NIBP
Measuring Technology Automatic oscillating measurement
Cuff Inflating <30s (0~300 mmHg, standard adult cuff)
Measuring Period AVE<40s
Mode Manual, Auto, STAT Manual, Auto
Measuring Interval AUTO 2 min ~ 4 hrs
PRR 30~250 (bpm)
Measuring Range:
Adult/Pediatric Mode SYS: 40~250 (mmHg)
DIA: 15~200 (mmHg)
Neonate Mode SYS: 40~135 (mmHg)
DIA: 15~100 (mmHg)
Accuracy:
Maximum Mean error ±5mmHg
Maximum SD 8mmHg
Resolution 1mmHg
Overpressure Protection Adult Mode: 280 (mmHg)
Neonate Mode: 150 (mmHg)
Adult Mode: 300 (mmHg)
Neonate Mode: 160 (mmHg)
Alarm Limit Setting SYS: 50~240 mmHg
DIA: 15~180 mmHg
TEMP
Range 25~50 (°C)
Accuracy ±0.2°C (25.0~34.9°C)
±0.1°C (35.0~39.9°C)
±0.2°C (40.0~44.9°C)
±0.3°C (45.0~50.0°C)
Display Res. 0.1°C
Alarm Limit Setting Upper limit 0~50°C
Lower limit 0~50°C
Channel 1 channels 2 channels
EtCO2
CO2 Measurement Range 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa 0~99mmHg
Accuracy 0~40 mmHg ±2mmHg
41~70 mmHg ±5% of reading
71~100 mmHg ±8% of reading
101~150 mmHg ±10% of reading
Above 80 breath per minute ± 12% of reading
±2mmHg (0~38mmHg)
39-99mmHg ±5% of reading +0.08% for every 1mmHg (abaove 38mmHg)
Sampling Rate 100Hz 50 ml/min
Initialization Time   30 seconds (typical), reaches ±5% steady-state accuracy within 3 minutes
Respiration Rate 2~150 bpm 0~150 breaths/min
Mode   Adult, neonate
Anesthetic Agents
Method   Infrared absorption
Gas Sorts   Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, O2 (optional Automatic Agent ID)
Measurement Range:
Halothane, Isoflurane   0~8.5%
Enflurane, Sevoflurane   0~10%
Desflurane   0~20%
CO2   0~10%
N2O   0~100%
O2   0~100%
Bias
Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane   ± (0.15 Vol% + 15% rel.)
CO2   ± (0.5 Vol% + 12% rel.)
N2O   ± (2 Vol% + 8% rel.)
O2   ±3 Vol%
Networking Wired Networking: Industry standard: IEEE 802.3 wired network
Connected beside number: Up to 16 bedside monitors RJ45 interface or RS232 serial port
 
Wireless Networking: Up to 100m indoors. Frequency Range: 2.412~2.484 GHz. Industry standard 802.11 b/g wireless Supports TCP/IP and UDP/IP Protocols
Industry standard 802.11 b/g wireless network
Power
Source External AC power or internal battery
AC Power 100~240VAC, 50/60Hz, 150VA
Battery Rechargeable Lead-Acid
Type: FB 1223 12v-2.3Ah
Operating time under the normal condition: 1 hour
Operating time after the first alarm of low battery: 10 minutes
Manufacturer: Pilot Battery Co.,Ltd.
Built-in & rechargeable lithium ion
Operating Time 1 hour 3+ hours
Environmental Specifications
Temperature Operating: 5~40°C
Storage: -10~45°C
Operating: 5~40°C
Storage: -20~65°C
Humidity range Operating: ≤80%
Storage: ≤80%